Thermovent 600 and 1200 Heat and Moisture Exchangers

Thermovent 600 and 1200 Heat and Moisture Exchangers

Thermovent 600 and 1200 are sterile, single use, heat and moisture exchangers for use with intubated patients using a ventilator or other positive pressure breathing system.

These devices are used to:

  • Preserve humidity and minimize damage to tracheal epithelial cells
  • Reduce heat loss through the bronchopulmonary tree
  • Help prevent thickened secretions and changes in lung function during long-term anesthesia or ventilatory support

Inquire about this product

Specifications / Ordering Information

Specifications

Thermovent 600

Thermovent 1200

Dead Space

11 mL and indicated for use in patients whose Tidal Volume is less than 600 mL including pediatric patients weighing more than 10 kg

Dead Space

32 mL and indicated for use in adult patients whose Tidal Volumes are greater than 600 mL and less than 1200 mL

Moisture Output

27mg/L H2O @ 20 breaths/min Tidal Volume 500 mL

Moisture Output: 24mg/L H2O @ 10 breaths/min Tidal Volume 1000 mL

Temperature Output

32.8°C @ 20 breaths/min Tidal Volume 500 mL

Temperature Output

33.0°C @ 10 breaths/min Tidal Volume 1000 mL

Resistance to flow

1.0 hPa (cmH2O) at gas flow rate 0.5 L/sec (30 L/min)

2.9 hPa (cmH2O) at gas flow rate 1.0 L/sec (60 L/min)

5.7 hPa (cmH2O) at gas flow rate 1.5 L/sec (90 L/min)

Resistance to flow

0.4 hPa (cmH2O) at gas flow rate 0.5 L/sec (30 L/min)

1.2 hPa (cmH2O) at gas flow rate 1.0 L/sec (60 L/min)

2.4 hPa (cmH2O) at gas flow rate 1.5 L/sec (90 L/min)

Compressible Volume

 11 mL (Dead space)

Compressible Volume

 32 mL (Dead space)

Thermovent Heat and Moisture Exchanging Filters

Reorder Code Pack Quantity
100/582/000 - Thermovent 1200 Sterile 20

Thermovent Heat and Moisture Exchanging Filters

Reorder Code Pack Quantity
580011 - Thermovent 600 Sterile 50